• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC. QUADRA ASSURA ICD; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC. QUADRA ASSURA ICD; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number CD3367-40QC
Device Problem Device Damaged Prior to Use (2284)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/08/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
During device preparation, the silicone on the header was damaged.
 
Manufacturer Narrative
Field experience of an anomalous rv df4 septum was verified in the laboratory.The rv df4 septum was found to be loose and partially detached from the set screw bore upon receipt in the laboratory.The cause of the loose septum could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUADRA ASSURA ICD
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.
645 almanor avenue
sunnyvale CA 94085
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.
645 almanor avenue
sunnyvale CA 94085
Manufacturer Contact
denise johnson
5050 nathan lane north
plymouth, MN 55442
6517564470
MDR Report Key6155794
MDR Text Key62232451
Report Number2938836-2016-14729
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/08/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/08/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/30/2018
Device Model NumberCD3367-40QC
Device Lot NumberP000028608
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/17/2016
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/17/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/28/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age72 YR
-
-