• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CYBERONICS, INC. PULSE GEN MODEL 106; GENERATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CYBERONICS, INC. PULSE GEN MODEL 106; GENERATOR Back to Search Results
Model Number 106
Device Problem Device Displays Incorrect Message (2591)
Patient Problem No Information (3190)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported by nurse that the following message was received during vns system check on (b)(6) 2016: 'pulse generator is disabled due to a burst watch dog time out.Not supplying stimulation'.It was reported that the generator was programmed with an autostim output current which was equal to the magnet output current 2ma.System diagnostics test confirmed that the impedance value is normal.The nurse was advised to adjust vns system¿s settings programing the magnet output current higher than the autostim output current.Review of manufacturing records confirmed all tests passed for the concerned generator prior to distribution.
 
Event Description
After further investigations, no data from the clinic visit dated (b)(6) 2016 could be found saved on the site's sd card.Information was received that the patient is now being titrated back up after the event and is now at 1.5ma output current and the signal frequency was dropped from 30 to 20 hz.The vns system impedance is normal (2607 ohms).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PULSE GEN MODEL 106
Type of Device
GENERATOR
Manufacturer (Section D)
CYBERONICS, INC.
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
CYBERONICS, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key6157479
MDR Text Key62380334
Report Number1644487-2016-02818
Device Sequence Number1
Product Code LYJ
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 01/24/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/09/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date09/22/2017
Device Model Number106
Device Lot Number4536
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Event Location Other
Date Manufacturer Received01/18/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/30/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age21 YR
-
-