(b)(4).Complaint sample was evaluated and the reported event was confirmed.Visual inspection of the inserter shows the distal threads are fractured.Fracture analysis of the part utilizing sem, suggests possible primary overload fracture starting at the lower side that propagated across to the higher fracture surface.Material was confirmed to be 455 stainless steel using an xrf scan.Only the last 1 threads were fractured on the inserter.Review of complaint history determined that no further action(s) is/are required.This part was manufactured, prior to a design change to decrease the brittleness of threads.Root cause was unable to be determined as the necessary information to adequately investigate the reported event was not provided.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|