• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HILL-ROM BATESVILLE CLINITRON AT*HOME; BED, AIR FLUIDIZED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HILL-ROM BATESVILLE CLINITRON AT*HOME; BED, AIR FLUIDIZED Back to Search Results
Model Number P0800010016
Device Problem Electrical /Electronic Property Problem (1198)
Patient Problem Pressure Sores (2326)
Event Date 11/17/2016
Event Type  Injury  
Manufacturer Narrative
The patient's existing pressure ulcer progressed to a stage 4, the mother was unsure of the prior stage, on one buttock.The patient receives wound care twice per month, with his parents providing continual in-home care.The wounds are being treated with alginate bandages and the patient is also taking oral antibiotics for the wound's infection.Development of pressure ulcers is multifactorial and cannot be only attributed to performance of the surface.Risk factors include protein-calorie malnutrition, microclimate (skin wetness caused by sweating or incontinence), diseases that reduce blood flow to the skin, such as arteriosclerosis, or diseases that reduce the sensation in the skin, such as paralysis or neuropathy.Position changes are key to pressure sore prevention and treatment.These changes need to be frequent, repositioning needs to avoid stress on the skin, and body positions need to minimize the risk of pressure on vulnerable areas.The hill-rom technician found the blower and electrical box needed to be replaced.The technician replaced the blower and electrical box to resolve the issue.Based on this information, no further action is required.Per the hill-rom user manual, poor fluidization; if fluidization is sluggish or uneven, notify your hill-rom representative.Fluidization is affected by the following: room temperature, humidity, the amount of materials, such as fluid, cells or cellular debris, which has escaped from the blood vessels and has been deposited in tissues or on tissue surfaces, restricted air circulation from blankets on the bed.It is unknown if the account performed any other preventative maintenance on this bed.
 
Event Description
Hill-rom received a report from the account alleging the patients wounds have gotten worse.The bed was located at the patients home.This report was filed in our complaint handling system as complaint #(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
BED, AIR FLUIDIZED
Manufacturer (Section D)
HILL-ROM BATESVILLE
1069 state route 46 east
batesville IN 47006
Manufacturer Contact
tony werner
1069 state route 46 east
batesville, IN 47006
8129312359
MDR Report Key6158106
MDR Text Key61887658
Report Number1824206-2016-00524
Device Sequence Number1
Product Code IOQ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K942184
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Reporter Occupation Patient
Type of Report Initial
Report Date 11/17/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/09/2016
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model NumberP0800010016
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received11/17/2016
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age32 YR
Patient Weight64
-
-