Product event summary: the full lead was returned, analyzed, and the inner insulation of the lead was extrinsically breached due to a cut.The outer insulation of the lead was extrinsically breached due to a cut, and the inner insulation and outer insulation of the lead and the rv conductor insulation was observed to have blood ingression.The analyst noted that during time of implant the overlay tube, outer insulation and rv conductor insulation was cut, because considerable blood is visible inside all insulation and rv conductor.Visual summary analysis of the lead indicated damage at implant.
|