Sterile processing department manager originally reported that the laparoscopic grasper was identified as broken during the procedure set-up protocol.Upon further investigation, it was determined that the laparoscopic grasper distal tip broke intraoperatively.Broken tip was retrieved immediately, removed from the patient, and procedure continued.There were no reported adverse outcomes related to this event.Since the original instrument was not provided for further investigation, six randomly selected laparoscopic graspers were pulled from the finished goods inventory in order to replicate the same findings as originally reported.All graspers were intact and functioned as originally intended.Laparoscopic grasper distal tip breakage may occur if use of excessive force was applied while the device was entered into the tissue.Other possible cause may be related to the incorrect position of the grasper while applying the end user¿s forces.Aesculap, the oem, was notified on december 09, 2016.
|