(b)(4).Date sent: 12/13/2016.Batch # n91w1z.Additional information was requested and the following was obtained: was the active blade tip broken off during the procedure? i do not know.Upon packaging of the instrument to be shipped i noticed the tip was broken off.I looked in the red biohazard bag that i had received the product in and was unable to locate the tip.I went to the account and spoke to the circulator and the scrub nurse neither of whom expressed concerns that the tip had broken off in the patient.The scrub explained that he actually squeezes the tip of the instrument when assembling and disassembling rather than closing the jaws with the handle.That being said, he thought he would have noticed for sure if something was off.The doctor was not in the or so i emailed him the findings.Did the broken blade tip fall into the patient? not that i know of.The staff did not think it did.Was the broken blade tip retrieved from the patient? no, but they don't think it ended up in there.Did this cause a change in the plan of care for the patient? at this point in time no.The device was returned with the distal tip of the blade broken off and not returned.The remaining blade portion was scratched, and with evidence of contact with metal in or out of the operative field.This blade tip portion may have broken off of the device during transport to our analysis site.The reported complaint was for: "they had received the error message relax pressure on blade which they addressed then the generator said tighten assembly".During functional testing on gen11 an alert screen was displayed.A probable cause of the device stop activating and display an alert screen is blade damage.The device was disassembled to inspect the internal components and no anomalies were found.Probable causes of blade damage, including breakage, are external contact during pre-op or general use, blade contact with other devices, staples or clips during the procedure or using any means other than the blade wrench to attach or detach the blade.Once minor blade damage has occurred, subsequent activations may increase damage severity and result in alert screens, such as ¿tighten assembly¿, ¿blade error detected¿ or "relaxed pressure on blade" followed by a ¿replace instrument¿ screen later in the procedure or "relaxed pressure on blade" followed by a ¿replace instrument¿ screen later in the procedure.The batch history record was reviewed and no defects, ncr¿s or protocols related to the complaint, were found during the manufacturing process.
|