• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AV-TEMECULA-CT XACT CAROTID STENT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AV-TEMECULA-CT XACT CAROTID STENT SYSTEM Back to Search Results
Catalog Number XRX04008T
Device Problem Inaccurate Delivery (2339)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/24/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.The device is expected to be returned for evaluation.It has not yet been received.A follow up report will be submitted with all relevant information.
 
Event Description
It was reported that the procedure was to treat a lesion in the moderately calcified, moderately tortuous, distal internal carotid artery.During deployment of the x-act freestyle tapered stent; the stent came out of sheath very fast and the stent moved over the lesion, from its intended placement during deployment.The stent deployed partially in the target lesion and partially in healthy tissue, not fully covering the target lesion.A second x-act stent was deployed overlapping the first stent to fully cover the lesion.There were no adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.
 
Manufacturer Narrative
(b)(4).Evaluation summary: visual inspection was performed on the returned device.The inaccurate delivery was not confirmed as the stent had already been deployed.It may be possible that anatomical conditions contributed to the reported inaccurate delivery; however, this could not be confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history of the reported lot revealed no other incidents.The investigation was unable to determine a conclusive cause for the reported inaccurate delivery.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XACT CAROTID STENT SYSTEM
Type of Device
CAROTID STENT SYSTEM
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, TEMECULA, CA USA REG# 2024168
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
connie speck
abbott vascular
26531 ynez road
temecula, CA 92591-4628
9519143996
MDR Report Key6178235
MDR Text Key62692526
Report Number2024168-2016-08901
Device Sequence Number1
Product Code NIM
Combination Product (y/n)N
Reporter Country CodeVM
PMA/PMN Number
P040038
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 02/01/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/15/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/31/2019
Device Catalogue NumberXRX04008T
Device Lot Number6041161
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/10/2017
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/31/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/01/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age64 YR
Patient Weight60
-
-