BOSTON SCIENTIFIC - COSTA RICA (HEREDIA) INTELLATIP MIFI¿ XP TEMPERATURE ABLATION CATHETER; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FLUTTER
|
Back to Search Results |
|
Model Number M004PM4790K20 |
Device Problem
No Device Output (1435)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 11/18/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Device evaluated by mfr: unit received in a decontamination pouch, overall visual did not identify failures or evidence that could be lost due to decontamination process.The returned device matches with upn and lot provided by the customer.There is a kink in the device approximately 14mm from the distal tip in the neutral position.The kink is between r1 and r2.The ring1 seal is split along the edge of the ring at the top of the kink¿s curve.There is dried body fluid along the sides of the ring.Ring 2 also has dried body fluid on the sides of the ring however the seals appear to be intact.The steering knob and the tension control knob functioned properly on both lock and unlock positions.The catheter was placed on the curve template and the device failed both the right and left curves tests; the tip did not reach the shaded area on the template in either direction.Electrical test was performed and the device was found to be within specifications.The ablation was verified by using the maestro generator 4000, and the device was found within specifications.The distal section was dissected.There is a large amount of body fluid in the interior of the sheath.Also, there is some body fluid under ring1 and ring2.The kevlar wrap is displaced and the center support is bent into a ¿v¿ shape.One of the steering wires is detached; there is evidence of solder on both the center support and the detached steering wire.The manufacturing batch record review confirms the device met all material, assembly and performance specifications.The root cause is not confirmed as there was no evidence of the alleged issue or any anomalies which could have contributed to the reported difficulty.(b)(4).
|
|
Event Description
|
It was reported that no signal occurred.After positioning the intellatip mifi¿ xp temperature ablation catheter in the patient, the cable was connected; however the electric potential was not displayed.Changing the cable did not resolve the issue.The issue was resolved by exchanging the catheter to another of the same device.No patient complications were reported and the patient's status is stable.However, returned device analysis revealed the ring seal was compromised.
|
|
Search Alerts/Recalls
|
|
|