• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PENUMBRA, INC. PENUMBRA SYSTEM ASPIRATION PUMP MAX 110V; JCX

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PENUMBRA, INC. PENUMBRA SYSTEM ASPIRATION PUMP MAX 110V; JCX Back to Search Results
Catalog Number PMX110
Device Problems Device Operates Differently Than Expected (2913); Pumping Problem (3016)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/19/2016
Event Type  malfunction  
Manufacturer Narrative
This device is available for return.A follow up mdr will be submitted upon completion of the device investigation.
 
Event Description
The patient was undergoing a thrombectomy procedure using a penumbra system aspiration pump max 110v (pump max).During the procedure, it was reported that the pump max was not working and was ¿acting funny¿.Therefore, the procedure was completed using a new pump max.It is unknown if there was an adverse effect to the patient.Please note that the manufacturer has requested additional information from the customer; however, no additional information has been obtained.
 
Manufacturer Narrative
Results: the vacuum gauge needle oscillated when the penumbra system aspiration pump max 110v (pump max) was generating vacuum.The accuracy of the vacuum gauge was within the specification.A small dent can be seen on the back of the pump max.Conclusions: evaluation of the device revealed the pump max vacuum gauge needle oscillated when the pump was generating vacuum.The accuracy of the vacuum gauge was tested, which calls for less than ±2 inhg total oscillation of the vacuum gauge when the pump is set to -25 inhg, and the pump max was found to be within specification.The pump max was able to power up repeatedly and produced consistent vacuum pressure.Penumbra pumps are visually and functionally inspected during incoming quality inspection.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENUMBRA SYSTEM ASPIRATION PUMP MAX 110V
Type of Device
JCX
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
Manufacturer Contact
veronica farris
one penumbra place
alameda, CA 94502
5107483200
MDR Report Key6178743
MDR Text Key62533112
Report Number3005168196-2016-01826
Device Sequence Number1
Product Code JCX
UDI-Device Identifier00814548012773
UDI-Public00814548012773
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K122756
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 11/19/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/15/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Catalogue NumberPMX110
Device Lot NumberF19378-18
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/27/2016
Is the Reporter a Health Professional? No
Date Manufacturer Received01/11/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/13/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-