A physician reported that they were receiving an error message stating "error establishing communication." troubleshooting was then performed on the physician's device, which included, checking the 9v battery and attempting to interrogate a known good generator.The communication issue persisted and the facility did not have another serial cable to interchange.A black serial cable was sent to the facility and reportedly resolved the communication issues.No patient therapy was affected by this malfunction.The facility then disposed of the white serial cable that was involved with the communication issues.The serial cable does not require return, as the failure mode is understood to be a failure of the serial cable associated with a disconnected wire connection.No other relevant information has been received to date.
|