Model Number MODEL 250
Device Problems Bent (1059); Defective Device (2588); Connection Problem (2900)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/22/2016
Event Type  malfunction  
Event Description
A physician's programmer was received by the manufacturer for analysis.During the analysis an intermittent power connection associated with the serial cable was identified.The cause for the anomaly is associated with a bent serial cable plug leaf spring that prevented electrical contact with the ac adapter barrel connector.Additionally, it was also identified that the main battery was defective and unable to power the handheld for over an hour.This is expected behavior considering the age of the handheld and main battery.No other anomalies were identified.No anomalies associated with flashcard software or databases were identified during and they performed according to functional specifications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGRAMMING SOFTWARE
Type of Device
PROGRAMMING COMPUTER
Manufacturer (Section D)
CYBERONICS - HOUSTON
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
CYBERONICS - HOUSTON
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key6183408
MDR Text Key62959897
Report Number1644487-2016-02908
Device Sequence Number1
Product Code LYJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial
Report Date 11/22/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/16/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 250
Device Lot Number1075839
Other Device ID NumberVERISON 8.1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/02/2016
Is the Reporter a Health Professional? No
Event Location Other
Date Manufacturer Received11/22/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/25/2009
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-