Catalog Number 15140
Device Problem Leak/Splash (1354)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/09/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The device sample was not returned for evaluation at the time of this report.
 
Event Description
The event is reported as: the customer alleges the mask was leaking after insertion.There was no patient harm reported.
 
Manufacturer Narrative
(b)(4).Device history record (dhr) review was performed and there were no issues found that could have contributed to the reported failure.Device was used and not in the original teleflex lma packaging.The connector was observed to be broken.The device could be inflated and deflated smoothly (holding air as intended).The device was tested in the water and there was no leak observed.The connector returned broken from the joint of the airway was most likely due to sudden abrupt force being applied when removing the connector.The reported defect of a leaking mask was unconfirmed as the device could be inflated and deflated as intended.
 
Event Description
The event is reported as: the customer alleges the mask was leaking after insertion.There was no patient harm reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LMA PROSEAL, REU, SIZE 4
Type of Device
LARYNGEAL MASK AIRWAY
Manufacturer (Section D)
TELEFLEX MEDICAL
athlone
Manufacturer (Section G)
THE LARYNGEAL MASK COMPANY
6 battery road #07-02
singapore 04990 9
SN   049909
Manufacturer Contact
warrenda peterson
3015 carrington mill blvd
morrisville, NC 27560
9193613959
MDR Report Key6184221
MDR Text Key62688309
Report Number9681900-2016-00060
Device Sequence Number1
Product Code CAE
Combination Product (y/n)N
Reporter Country CodeAS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/12/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/17/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Catalogue Number15140
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/20/2016
Is the Reporter a Health Professional? No
Date Manufacturer Received02/06/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-