Further investigation of the customer issue included a review of the complaint text, a search for similar complaints, and a review of labeling.Return parts were not available.Further investigation of the customer issue included a review of the complaint text, a search for similar complaints, and a review of labeling.Review of the submitted data showed the cell-dyn ruby sent the correct data for (b)(4) to the lis system.Based on these results, the system was performing as expected.Tracking and trending did not identify an adverse trend.Labeling was reviewed and found to be adequate.Based on the available information no product deficiency of the cell-dyn ruby analyzer, list number 08h67 was identified.Udi: (b)(4) lot/serial number was manufactured prior to udi compliance date; therefore, only a di is provided.
|