ABBOTT DIAGNOSTICS DIVISION CELL-DYN RUBY ANALYZER; AUTOMATED HEMATOLOGY ANALYZER Back to Search Results
Catalog Number 08H67-01
Device Problem Patient Data Problem (3197)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
A follow-up report will be submitted when the evaluation is complete.
 
Event Description
The customer observed a sample id mismatch while using the cell-dyn ruby analyzer.The customer indicated that the technician noticed the error and did not release the results associated with the incorrect id.The customer stated that 3 different specimens were impacted.The following example was provided: results for sid (b)(6) were charted from the lis for sid (b)(6).The results were corrected in the lis system.No impact to patient management was reported.
 
Manufacturer Narrative
Further investigation of the customer issue included a review of the complaint text, a search for similar complaints, and a review of labeling.Return parts were not available.Further investigation of the customer issue included a review of the complaint text, a search for similar complaints, and a review of labeling.Review of the submitted data showed the cell-dyn ruby sent the correct data for (b)(4) to the lis system.Based on these results, the system was performing as expected.Tracking and trending did not identify an adverse trend.Labeling was reviewed and found to be adequate.Based on the available information no product deficiency of the cell-dyn ruby analyzer, list number 08h67 was identified.Udi: (b)(4) lot/serial number was manufactured prior to udi compliance date; therefore, only a di is provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CELL-DYN RUBY ANALYZER
Type of Device
AUTOMATED HEMATOLOGY ANALYZER
Manufacturer (Section D)
ABBOTT DIAGNOSTICS DIVISION
4551 great america parkway
santa clara CA 95054
Manufacturer (Section G)
ABBOTT DIAGNOSTICS DIVISION
4551 great america parkway
santa clara CA 95054
Manufacturer Contact
noemi romero-kondos, rn bsn
100 abbott park road
dept. 09b9, lccp1-3
abbott park, IL 60064-3537
224667-512
MDR Report Key6186234
MDR Text Key62793469
Report Number2919069-2016-02314
Device Sequence Number1
Product Code GKZ
UDI-Device Identifier00380740017170
UDI-Public00380740017170
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K061667
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 01/30/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/19/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number08H67-01
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/11/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/16/2006
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-