Catalog Number 254500547 |
Device Problem
Break (1069)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 12/13/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.This complaint is still under investigation.Depuy will notify the fda of the results of this investigation once it has been completed.
|
|
Event Description
|
The surgeon was inserting the articulation surface to trial and the articulation surface broke.
|
|
Manufacturer Narrative
|
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.This complaint is still under investigation.Depuy will notify the fda of the results of this investigation once it has been completed.
|
|
Manufacturer Narrative
|
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.This complaint is still under investigation.Depuy will notify the fda of the results of this investigation once it has been completed.
|
|
Manufacturer Narrative
|
Examination of the returned device confirms the reported event of trial breakage.Hhe/qrb (quality review board) 103045639 recommended a device correction which was initiated on june 12, 2015.Capa-(b)(4) determined the likely root cause to be related to misuse, and requires mandatory sales training on proper use and application of the device by depuy sales consultants.Capa-(b)(4) will also monitor the mandatory field training.The need for further corrective action was not indicated.Continue to monitor per post market surveillance sep-419.Depuy considers the investigation closed at this time.Should the additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.
|
|