Catalog Number 397002-001 |
Device Problem
Device Displays Incorrect Message (2591)
|
Patient Problems
No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
|
Event Date 12/13/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
This alleged failure mode poses a low risk to the patient because although the companion 2 driver exhibited a malfunction alarm, the companion 2 driver continued to perform its life-sustaining functions.The companion 2 driver will be returned to syncardia for evaluation.The results of the evaluation will be provided in a follow-up mdr.(b)(4) initial.
|
|
Event Description
|
The customer, a syncardia certified hospital, reported that the companion 2 driver exhibited a system malfunction alarm.The patient was subsequently switched to the backup companion 2 driver.There was no reported adverse patient impact.
|
|
Manufacturer Narrative
|
The companion 2 driver was returned to syncardia for evaluation.During investigation testing, the system malfunction alarm was reproduced.Analysis of data collected from the pressure sensor printed circuit assembly (pca) confirmed a malfunction of the pressure transducer at m7 on the pressure sensor pca.This malfunction caused the driver to perceive an internal filter block and the system malfunction alarm was triggered.This issue will continue to be monitored and trended as part of the customer experience process.Syncardia has completed its evaluation of this complaint and is closing this file.(b)(4).
|
|
Event Description
|
The customer, a syncardia certified hospital, reported that the companion 2 driver exhibited a system malfunction alarm.The patient was subsequently switched to the backup companion 2 driver.There was no reported adverse patient impact.
|
|