• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: 3M HEALTH CARE 3M RANGER(TM) HIGH FLOW BLOOD/FLUID DISPOSABLE SET; HIGH FLOW DISPOSABLE TUBING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

3M HEALTH CARE 3M RANGER(TM) HIGH FLOW BLOOD/FLUID DISPOSABLE SET; HIGH FLOW DISPOSABLE TUBING Back to Search Results
Model Number N/A
Device Problem Appropriate Term/Code Not Available (3191)
Patient Problem Not Applicable (3189)
Event Date 12/05/2016
Event Type  malfunction  
Manufacturer Narrative
The product was not returned for evaluation as of the date of this report.3m implemented a new dehp free material to enhance the material properties on all 3m ranger(tm) fluid warming disposable products in 2013.Subsequent to this change 3m received a higher trend of complaints applicable to leaks within selected areas of the disposable tubing connections.3m implemented a solvent improvement process change in late 2015 to address this type of event.This improvement was to the solvent bond process to reduced leaks on all of the following areas; y-connector, drip chamber, bubble trap and tube fitting of the high flow disposable set.The change was implemented to increase strength and to minimize leaks.The reported lot in this event did include the solvent process change.3m continues to monitor their complaints to confirm the effectiveness of the change made and is also monitoring trends of complaints subsequent to the change.End of report.
 
Event Description
A hospital employee reported that a 3m(tm) ranger(tm) high flow blood/fluid disposable set leaked at the y connector.There was no reported patient injury.No information was provided as to when the leak occurred nor the type of fluid being used.
 
Manufacturer Narrative
Updated the report to include initial reporter name, e-mail and phone number and also full address of hospital.Also indicated more information about the use of product and when leak occurred.End of report.
 
Event Description
The leak occurred before it was attached to the patient and during priming.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3M RANGER(TM) HIGH FLOW BLOOD/FLUID DISPOSABLE SET
Type of Device
HIGH FLOW DISPOSABLE TUBING
Manufacturer (Section D)
3M HEALTH CARE
2510 conway ave
st. paul MN 55144
Manufacturer (Section G)
NYPRO PRECISION ASSEMBLIES
sor juana ines de la cruz
cd. industrial
tijuana, tijuana baja california 22444
MX   22444
Manufacturer Contact
linda johnsen
3m center, building 275-5w-06
2510 conway ave
st. paul, MN 55144
6517374376
MDR Report Key6187433
MDR Text Key62791049
Report Number2110898-2016-00113
Device Sequence Number1
Product Code BSB
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K973741
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/21/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/19/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/28/2018
Device Model NumberN/A
Device Catalogue Number24355
Device Lot NumberHX7646
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received12/21/2016
Was Device Evaluated by Manufacturer? No
Date Device Manufactured12/01/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-