SIEMENS HEALTHCARE, BUSINESS AREA ADVANCED THERAPY ARTIS ZEEGO; SYSTEM, X-RAY, ANGIOGRAPHIC
|
Back to Search Results |
|
Model Number 10280959 |
Device Problem
Self-Activation or Keying (1557)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 11/22/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
Siemens is conducting a thorough investigation of the reported events.As this event is under investigation, a root cause has not yet been determined.A supplement report will be filed upon completion of the investigation.
|
|
Event Description
|
It was reported to siemens that a malfunction occurred while operating the artis zeego system.During a patient procedure, it was reported that the system dropped into bypass fluoro.A system restart was attempted, however, the intubated patient was safely removed from the system and transferred to an alternative system where the procedure was completed.We are unaware of any impact to the state of health of the patient involved.
|
|
Manufacturer Narrative
|
Siemens has completed an investigation of the reported event.During the investigation, a defective ethernet-switch was identified.The entire ias pc (image acquisition system) as well as the ethernet-switch were exchanged on site.Following the exchange of components, the system works as specified and the issue did not recur.The manufacturer is not considering further actions resulting from this event.
|
|