• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SIEMENS HEALTHCARE, BUSINESS AREA ADVANCED THERAPY ARTIS ZEEGO; SYSTEM, X-RAY, ANGIOGRAPHIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIEMENS HEALTHCARE, BUSINESS AREA ADVANCED THERAPY ARTIS ZEEGO; SYSTEM, X-RAY, ANGIOGRAPHIC Back to Search Results
Model Number 10280959
Device Problem Self-Activation or Keying (1557)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/22/2016
Event Type  malfunction  
Manufacturer Narrative
Siemens is conducting a thorough investigation of the reported events.As this event is under investigation, a root cause has not yet been determined.A supplement report will be filed upon completion of the investigation.
 
Event Description
It was reported to siemens that a malfunction occurred while operating the artis zeego system.During a patient procedure, it was reported that the system dropped into bypass fluoro.A system restart was attempted, however, the intubated patient was safely removed from the system and transferred to an alternative system where the procedure was completed.We are unaware of any impact to the state of health of the patient involved.
 
Manufacturer Narrative
Siemens has completed an investigation of the reported event.During the investigation, a defective ethernet-switch was identified.The entire ias pc (image acquisition system) as well as the ethernet-switch were exchanged on site.Following the exchange of components, the system works as specified and the issue did not recur.The manufacturer is not considering further actions resulting from this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARTIS ZEEGO
Type of Device
SYSTEM, X-RAY, ANGIOGRAPHIC
Manufacturer (Section D)
SIEMENS HEALTHCARE, BUSINESS AREA ADVANCED THERAPY
siemensstrasse 1
forcheim, 91301
GM  91301
Manufacturer (Section G)
SIEMENS HEALTHCARE GMBH
siemenstrasse 1
forcheim, 91301
GM   91301
Manufacturer Contact
meredith adams
40 liberty blvd.
65-1a
malvern, PA 19355
6104486461
MDR Report Key6188855
MDR Text Key63199087
Report Number2240869-2016-62040
Device Sequence Number1
Product Code IZI
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K141574
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Remedial Action Inspection
Type of Report Initial,Followup
Report Date 11/22/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/20/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model Number10280959
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received12/05/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/05/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-