• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LTD. - HOUSTON CENTURION ULTRAVIT PROBE PROCEDURE PAK; GENERAL SURGERY TRAY (KIT)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LTD. - HOUSTON CENTURION ULTRAVIT PROBE PROCEDURE PAK; GENERAL SURGERY TRAY (KIT) Back to Search Results
Catalog Number 8065752134
Device Problem Kinked (1339)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/07/2016
Event Type  malfunction  
Manufacturer Narrative
Investigation, including root cause analysis, is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.(b)(4).
 
Event Description
A customer reported kinked tubing noted during an eye surgery.The product was replaced and the procedure was completed without harm to the patient.A product sample was requested for evaluation.
 
Manufacturer Narrative
A device history record of the reported lot shows that the order was built to specification.A sample has not been returned for this complaint; therefore, visual inspection and functional testing could not be performed.The root cause of the customer's complaint could not be established as a sample has not been received.Without a sample, it is not possible to isolate the root cause.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTURION ULTRAVIT PROBE PROCEDURE PAK
Type of Device
GENERAL SURGERY TRAY (KIT)
Manufacturer (Section D)
ALCON RESEARCH, LTD. - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer (Section G)
ALCON RESEARCH, LTD. - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer Contact
rita lopez
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8175514846
MDR Report Key6189279
MDR Text Key63213461
Report Number1644019-2016-01554
Device Sequence Number1
Product Code LRO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K880961
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other
Reporter Occupation Other
Remedial Action Other
Type of Report Initial,Followup
Report Date 02/13/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/20/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/30/2018
Device Catalogue Number8065752134
Device Lot Number1885699H
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received02/08/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/16/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-