Quality evaluation of the returned plate found both dimensions that affect mating of rivet with plate to be within tolerance.All plates of lot 8032ps were 100% inspected during manufacturing process for these two dimensions.Visual review/inspection of the simplicity anterior cervical plate 1-level 20mm (acp120) returned showed no obvious signs of damage to the plate that would cause the rivet to fail.The surgeon had indicated that screw may have been a little proud which may be a result of over angulation of the screw.Attempting to lock/rotate the locking rivets over a proud screw could cause the fracturing of the rivet due to a ramping effect.With the information/product provided, a proud screw is believed to be the root cause of failure.Review of manufacturing history records found a total of (b)(4) pieces of lot 8032ps were released for distribution in january 2016, with no deviation or anomalies.Two-year complaint history review did not reveal any complaints of locking rivets breaking during implantation.As the root cause is attributed to technique, and there is not a trend for reports of this nature, the need for corrective action was not indicated.
|