MAUDE Adverse Event Report: BIOMET ORTHOPEDICS 52MM HEAD HIP MOLD W/INSERT; PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET ORTHOPEDICS 52MM HEAD HIP MOLD W/INSERT; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Device Problem Device Contaminated During Manufacture or Shipping (2969)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/22/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that when the packaging was opened, there was a hair inside the sterile packaging.There was no back up device of the same size available, so the surgeon implanted a smaller size.
 
Manufacturer Narrative
Visual inspection of the returned packaging of the product confirmed that there is a hair-like structure inside the sterile packaging.Analysis of debris identified that it is stiff like a metallic wire or bristle and is consistent with 300-series stainless steel.Dhr was reviewed and no discrepancies relevant to the reported event were found.Review of the complaint history determined that no further action is required as no trends were identified.The root cause of the reported event is attributed to a trained operator error.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Manufacturer Narrative
If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
52MM HEAD HIP MOLD W/INSERT
Type of Device
PROSTHESIS, HIP
Manufacturer (Section D)
BIOMET ORTHOPEDICS
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
BIOMET ORTHOPEDICS
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key6191392
MDR Text Key62910028
Report Number0001825034-2016-05313
Device Sequence Number1
Product Code KWY
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
PK080979
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign,health pr
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 02/03/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/20/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/16/2021
Device Model NumberN/A
Device Catalogue Number431182
Device Lot Number591240
Other Device ID NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/23/2017
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/02/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/16/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
-
-