The irc0lxo2 concentrator was returned for evaluation, and subsequent testing did not verify the complaint.Per the evaluation: visual inspection indicated that the hour meter read 186.35 hours.The concentrators exterior found minor cosmetic flaws but no structural damage.Both the power cable & plug were inspected and found to be undamaged.Inspection of the inlet filter found two large dark spots on the filter trap.Functional testing: at power up, the concentrator appropriately gave a start-up alarm.The flow meter was set to 0 lpm and the concentrator gave a flow obstruction alarm.The flow meter was opened & closed several times and the alarm sounded each time the meter was closed, thus, the reported issue of the concentrator not alarming for flow obstruction could not be confirmed.Next, the power plug was pulled and the concentrator appropriately alarmed for sudden power loss.In summary, basic alarm functionality was confirmed.A basic performance test at full scale of the flow meter was conducted.The concentrator was allowed to run for 30 minutes continuously.No alarms were observed during this period.After 30 minutes output pressure and o2 purity were measured at 9.4 psi and 95.5% for a flow rate of 9.9 lmp.Based on the maximum output pressure and acceptable o2 purity listed in the service manual the concentrator functioned properly at full scale of the flow meter.The 6 lpm setting was tested and output pressure and o2 purity were measured at 9.5 psi and 95.5% respectively.The 4 lpm setting was also tested with output pressure and o2 purity measuring at 9.5 psi and 95.6% respectively.The concentrator functioned properly, therefore, the reported issue of the concentrator giving no output when powered on could not be confirmed.Conclusion: in its "as received" condition, the complaint was not confirmed.The irc10lxo2's flow obstruction alarm was observed to operate properly.Likewise, output pressure and o2 purity at 4 lpm, 6 lpm and full scale of the flow meter were found to satisfy typical normal operating parameters, thus indicating proper oxygen output.Thus, no malfunction could be confirmed.
|