MERIT MEDICAL SYSTEMS, INC MERIT CUSTOM KIT Back to Search Results
Catalog Number K12-MS2666A
Device Problem Device Contamination with Chemical or Other Material (2944)
Patient Problem No Patient Involvement (2645)
Event Date 11/09/2016
Event Type  malfunction  
Manufacturer Narrative
One unit was returned for evaluation.The unit was visually inspected.The complaint is confirmed.The root cause is attributed to the manufacturing process.The device history record was reviewed and no exception documents were found.The complaint database was reviewed and no related complaints for this lot number were found.Corrective actions are in process.
 
Event Description
The distributor alleged a foreign object was inside the fluid path of the syringe.This was identified during their initial inspection of received product.The device was not sent to a user facility.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MERIT CUSTOM KIT
Type of Device
CUSTOM KIT
Manufacturer (Section D)
MERIT MEDICAL SYSTEMS, INC
1600 west merit parkway
south jordan UT 84095
Manufacturer (Section G)
MERIT MEDICAL SYSTEMS, INC
1600 west merit parkway
south jordan UT 84095
Manufacturer Contact
casey hughes, ms, cqe
1600 west merit parkway
south jordan, UT 84095
8013164932
MDR Report Key6199642
MDR Text Key63133068
Report Number1721504-2016-00248
Device Sequence Number1
Product Code DQO
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K913682
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Other
Type of Report Initial
Report Date 11/09/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/22/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/31/2019
Device Catalogue NumberK12-MS2666A
Device Lot NumberH959797
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/29/2016
Date Manufacturer Received12/12/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/20/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-