Model Number 4415
Device Problem Failure to Prime (1492)
Patient Problem No Patient Involvement (2645)
Event Date 11/29/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The issue was discovered when the unit was tested and it did not involve any procedure.
 
Event Description
It was reported that during the use of the device for a non-clinical activity, the unit was not priming.There was no patient involvement.
 
Manufacturer Narrative
(b)(4).The reported issue was verified by the field service representative (fsr).After troubleshooting, a malfunctioning pressure switch was found to be the source of the problem.Thus, the pressure sensor was replaced.The unit operated to the manufacturer's specifications.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SARNS TEMP. CONTROL AND MONITOR SYSTEM (TCM II)
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS-SARNS TCM II HEATING AND COOLING
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORP.
6200 jackson road
ann arbor MI 48103
Manufacturer Contact
eileen dorsey
6200 jackson road
ann arbor, MI 48103
7347416074
MDR Report Key6202517
MDR Text Key63611830
Report Number1828100-2016-00805
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K883603
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/24/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/23/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number4415
Device Catalogue Number4415
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/17/2017
Is the Reporter a Health Professional? No
Date Manufacturer Received03/01/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/27/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-