• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS, INC. 45 DEG ANGLE DRILL; HIP INSTRUMENT/TRIAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS, INC. 45 DEG ANGLE DRILL; HIP INSTRUMENT/TRIAL Back to Search Results
Catalog Number 236683000
Device Problem Break (1069)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/14/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).This complaint is still under investigation.Depuy will notify the fda of the results of this investigation once it has been completed.
 
Event Description
After use, the drill bit, a small spring and the encapture for the drill bit fell out of the angle drill.
 
Manufacturer Narrative
Examination of the returned instrument confirmed the complaint.The root cause appears to be attributed to misuse; the drill bit may have been pulled out instead of turning to remove as intended.A complaint database search finds no other reported incidents against the provided product and lot combination.The investigation could not verify or identify any product contribution to the reported event with the information provided.Based on the investigation, the need for corrective action is not indicated.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
45 DEG ANGLE DRILL
Type of Device
HIP INSTRUMENT/TRIAL
Manufacturer (Section D)
DEPUY ORTHOPAEDICS, INC.
700 orthopaedic drive
warsaw IN 46582
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC.
700 orthopaedic drive
warsaw IN 46582
Manufacturer Contact
chad gibson
700 orthopaedic drive
warsaw, IN 46581
5743725905
MDR Report Key6204571
MDR Text Key63261751
Report Number1818910-2016-33898
Device Sequence Number1
Product Code HTW
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,health profession
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 12/14/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/27/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number236683000
Device Lot NumberSO2022815
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/19/2016
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/09/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/22/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-