Model Number DIMENSION VISTA 1500 INTERNATIONAL |
Device Problem
Low Test Results (2458)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 11/14/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
A siemens healthcare diagnostics inc customer service engineer (cse) was dispatched to the account.No hardware issues were identified.Siemens headquarters support center analysis of the instrument data indicate that the cause for the discordant low glucose csf results is unknown.Although two separate sample ids were stated to contain atypical low results, it was noted that both had originated from the same patient.In each instance the sample was repeated from a new aliquot well and recovered considerably higher and within the reference range.Additionally, all results were based on the same calibration curve and used the same reagent flex well suggesting that this issue was neither due to calibration bias nor a problem with the reagent material.The associated error logs for this event show multiple aliquot related errors for bad sample detect and clog detected.The instrument is performing within specifications.No further evaluation of the device is required.
|
|
Event Description
|
Discordant, low glucose (glu) results were obtained on one patient's spinal fluid (csf) samples on the dimension vista 1500 instrument.The discordant results were not reported to the physician.The samples were repeated and correct results were reported to the physician.There are no reports of patient intervention or adverse health consequences due to the discordant, low glucose csf results.
|
|
Search Alerts/Recalls
|