The unit was received with critical pump error and broken force sensor in the format history file due to moisture damage on force sensor.Unable to perform displacement test, rewind test, prime/seating test, basic occlusion test, force sensor test, occlusion test, sleep current measurement, active current measurement and self test due to critical pump errors.The following physical damage was noted: scratched casing and minor scratches on lcd window, scratches on display window cover, pillowing keypad overlay, cracked retainer, missing end cap address label, corrosion on all electronic assemblies and moisture damage on the motor home switch.The insulin pump involved in this event is the 640g insulin infusion pump, which is not marketed in the united states.However, the device is similar to the paradigm real-time insulin infusion pump, which is marketed in the united states.(b)(4).
|