Model Number B-20260
Device Problem Fluid/Blood Leak (1250)
Patient Problems Regurgitation (2259); Patient Problem/Medical Problem (2688)
Event Date 07/08/2013
Event Type  malfunction  
Manufacturer Narrative
Device labeling addresses the reported events as follows: precautions: care must be taken during band adjustment to avoid puncturing the tubing that connects the access port and band, as this will cause leakage and deflation of the inflatable section.Failure to create a stable, smooth path for the access port tubing, without sharp turns or bends, can result in tubing breaks and leakage.In order to avoid incorrect placement, the port should be placed lateral to the trocar opening.A pocket must be created for the port so that it is placed far enough from the trocar path to avoid abrupt kinking of the tubing.When adjusting band volume, use of an inappropriate needle may cause access port leakage and require reoperation to replace the port.Use only lap-band ap® system access port needles.Do not use standard hypodermic needles, as these may cause leaks.Adverse events: it is important to discuss all possible complications and adverse events with your patient.Complications which may result from the use of this product include the risks associated with the medications and methods utilized in the surgical procedure, the risks associated with any surgical procedure and the patient's degree of intolerance to any foreign object implanted in the body.Ulceration, gastritis, gastroesophageal reflux, heartburn, gas bloat, dysphagia, dehydration, constipation, and weight regain have been reported after gastric restriction procedures.Deflation of the band may occur due to leakage from the band, the port or the connecting tubing.Warnings: patients should be advised that the lap-band ap system is a long-term implant.Explant and replacement surgery may be indicated at any time.Medical management of adverse reactions may include explantation.Revision surgery for explantation and replacement may also be indicated to achieve patient satisfaction.
 
Event Description
Reported as: the patient with lap-band system had "reflux and band in a deflated state for the last 3 years.Now all weight lost returned and wanted to inflate the band." physician arranged for a radiological fill but the fill restriction was very short lasting.Later detected a leak in the tubing joint on the port side.Physician had the tubing trimmed and reconnected.Device remains implanted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LAP-BAND AP SMALL WITH ACCESS PORT II SYSTEM
Type of Device
ADJUSTABLE GASTRIC BAND
Manufacturer (Section D)
APOLLO ENDOSURGERY, INC.
1120 s capitol of texas hwy
bldg 1, ste 300
austin TX 78746
Manufacturer (Section G)
ALLERGAN
global park free zone
900 global park
la aurora de heredia
Manufacturer Contact
laura leboeuf
1120 s capital of texas hwy
bldg 1, ste 300
austin, TX 78746
5122795141
MDR Report Key6209281
MDR Text Key63476480
Report Number3006722112-2016-00404
Device Sequence Number1
Product Code LTI
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
P000008
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 11/29/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/28/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Expiration Date10/23/2010
Device Model NumberB-20260
Device Catalogue NumberB-20260
Device Lot Number1670283
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/29/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/27/2008
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age48 YR
Patient Weight112
-
-