Catalog Number 2C6256 |
Device Problem
Disconnection (1171)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 12/05/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
Potential lot numbers are dr16g09040 and dr16g30020.Should additional relevant information become available, a supplemental report will be submitted.
|
|
Event Description
|
It was reported that a clearlink system continu-flo solution set had disconnected.This occurred during patient infusion.It was stated that the disconnection happened during a pre-operative procedure.They completed the procedure using a new set.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
|
|
Manufacturer Narrative
|
Mfg date: (b)(4) was manufactured 07-11-2016; (b)(4) was manufactured 07-20-2016.The device was received for evaluation.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.Functional testing was performed with no leaks identified.The device was found to operate per specification.A batch review was conducted for each potentially associated lot and there were no deviations found related to this reported condition during the manufacture of either lot.Should additional relevant information become available, a supplemental report will be submitted.
|
|