• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER MEDICAL-KALAMAZOO CAPITAL TEMP PUMP PROFESSIONAL; PACK, HOT OR COLD, WATER CIRCULATING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER MEDICAL-KALAMAZOO CAPITAL TEMP PUMP PROFESSIONAL; PACK, HOT OR COLD, WATER CIRCULATING Back to Search Results
Catalog Number TP700C
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/30/2016
Event Type  malfunction  
Event Description
It was reported that the pump was experiencing failures of the backup limit thermostat test.No patient was affected and no adverse consequence or clinically relevant delay in treatment was reported.
 
Manufacturer Narrative
The issue was resolved for the customer by replacing the unit.
 
Event Description
It was reported that the pump was experiencing failures of the backup limit thermostat test.No patient was affected and no adverse consequence or clinically relevant delay in treatment was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAPITAL TEMP PUMP PROFESSIONAL
Type of Device
PACK, HOT OR COLD, WATER CIRCULATING
Manufacturer (Section D)
STRYKER MEDICAL-KALAMAZOO
3800 east centre avenue
portage MI 49002
Manufacturer (Section G)
STRYKER MEDICAL-KALAMAZOO
3800 east centre avenue
portage MI 49002
Manufacturer Contact
chanda burghard
3800 east centre avenue
portage, MI 49002
2693292100
MDR Report Key6213985
MDR Text Key63820144
Report Number0001831750-2016-00426
Device Sequence Number1
Product Code ILO
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 05/09/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/29/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Catalogue NumberTP700C
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received11/30/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/01/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-