Concomitant medical products: product id 37603, serial# (b)(4), implanted: (b)(6) 2016, product type: implantable neurostimulator.Product id 3387s-40, lot# va1ag0z, implanted: (b)(6) 2016, product type: lead.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
Additional information was received from a manufacturer representative (rep) regarding a patient who was implanted with a neurostimulator for dystonia and movement disorders.It was reported that it was determined that the lead was damaged at some point.After testing impedances in several areas, it was determined that the issue was the lead.The resolution plan was to program around the affected contacts.Information was previously reported about this event was sent using manufacturer report # 3007566237-2016-04246, additional information will also be sent under this manufacturers report number".
|