Catalog Number 254500545
Device Problem Break (1069)
Patient Problem No Information (3190)
Event Date 01/02/2017
Event Type  malfunction  
Manufacturer Narrative
This complaint is still under investigation.Depuy will notify the fda of the results of this investigation once it has been completed.(b)(4).
 
Event Description
The articulating trial broke when removing after trialing.
 
Manufacturer Narrative
This complaint is still under investigation.Depuy will notify the fda of the results of this investigation once it has been completed.
 
Manufacturer Narrative
Examination of the returned device confirms the reported event of post breakage.(b)(4) recommended a device correction which was initiated on june 12, 2015.Capa-(b)(4) determined the likely root cause to be related to misuse, and requires mandatory sales training on proper use and application of the device by depuy sales consultants.Capa-(b)(4) will also monitor the mandatory field training.The need for further corrective action was not indicated.Continue to monitor per post market surveillance sep-419.Depuy considers the investigation closed at this time.Should the additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE RP PS ARTIC SURF SZ5
Type of Device
KNEE INSTRUMENTS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic drive
warsaw IN 46582
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic drive
warsaw IN 46582
Manufacturer Contact
chad gibson
700 orthopaedic drive
warsaw, IN 46581
5743725905
MDR Report Key6222892
MDR Text Key63867191
Report Number1818910-2017-10168
Device Sequence Number1
Product Code NJL
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K140881
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,health profession
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 01/02/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/04/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number254500545
Device Lot NumberMVMBSY930
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/10/2017
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/02/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/06/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age73 YR
-
-