• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. ACTIVA; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN TREMOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. ACTIVA; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN TREMOR Back to Search Results
Model Number 37601
Device Problem Low impedance (2285)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/09/2016
Event Type  malfunction  
Manufacturer Narrative
Section information references the main component of the system, other applicable components are: product id 3387s-40 lot# va0k227 serial# implanted: (b)(6) 2014 explanted: product type lead product id 3387s-40 lot# va0k227 serial# implanted: (b)(6) 2014 explanted: product type lead.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from a manufacturer representative (rep) regarding a patient who was implanted with a neurostimulator for obsessive compulsive disorder (ocd) and movement disorders.It was reported that an impedance measurement was taken prior to an eeg being performed on the patient.The impedance measurement resulted in electrodes 8 <(>&<)> 9 having values of 26 ohms; a short was reported.It was confirmed that the patient was programmed on c and 8- so the battery and therapy were not affected.Follow-up with the rep indicated that no troubleshooting was performed to determine the cause of the short as the patient was at a mri facility to receive a brain scan.It was also stated that there were no actions on the rep's part to resolve the issue as the patient was not programmed on that particular contact and was receiving therapy without any issues.It is unknown if the impedance issue has resolved.No symptoms were reported.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACTIVA
Type of Device
STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN TREMOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
diane wolf
7000 central avenue ne rcw215
minneapolis, MN 55432
7635263987
MDR Report Key6222982
MDR Text Key63870096
Report Number3004209178-2017-00208
Device Sequence Number1
Product Code MFR
UDI-Device Identifier00643169529786
UDI-Public00643169529786
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
H050003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/04/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/04/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/14/2016
Device Model Number37601
Device Catalogue Number37601
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received12/09/2016
Date Device Manufactured06/26/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age26 YR
-
-