Model Number 10-80-00
Device Problems Defective Alarm (1014); Pumping Stopped (1503); Device Displays Incorrect Message (2591)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/14/2016
Event Type  No Answer Provided  
Manufacturer Narrative
Patient information was not provided.Sorin group (b)(4) manufactures the s5 roller pump.The incident occurred in (b)(6).This medwatch report is being filed on behalf of sorin group (b)(4).A sorin group field service representative was dispatched to the facility to investigate.The service representative tested the device extensively and was unable to reproduce or confirm the reported issue.The shaft angle encoder was replaced as a precaution and subsequent testing did not identify any issues.The unit was returned to service.As the issue could not be reproduced, a root cause could not be determined and no corrective actions were identified.A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.Sorin group (b)(4) will continue to monitor for trends related to this type of issue.Evaluated on site by sorin service rep.
 
Event Description
Sorin group (b)(4) received a report that the s5 roller pump failed and displayed an alarm message during a procedure.The pump was reset to clear the alarm and continue the case.There was no report of patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S5 ROLLER PUMP
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
SORIN GROUP DEUTSCHLAND
lindberghstr. 25
munich, 80939
GM  80939
Manufacturer (Section G)
SORIN GROUP DEUTSCHLAND
lindberghstr. 25
munich, 80939
GM   80939
Manufacturer Contact
joan ceasar
14401 w. 65th way
arvada, CO 80004
2812287260
MDR Report Key6227456
MDR Text Key64018518
Report Number9611109-2016-00952
Device Sequence Number1
Product Code DWB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K071318
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Biomedical Engineer
Remedial Action Other
Type of Report Initial
Report Date 12/14/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/05/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number10-80-00
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/14/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/27/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-