• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) TENDRIL STS; PERMANENT PACEMAKER ELECTRODE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) TENDRIL STS; PERMANENT PACEMAKER ELECTRODE Back to Search Results
Model Number 2088TC/52
Device Problems Bent (1059); Kinked (1339)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/21/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
The lead was damaged as it was being inserted into the sheath.The tip of the lead would not advance past the valve.We switched out the sheath, but the lead was already too damaged to be used and a new lead placed successfully.The patient was stable post procedure.
 
Manufacturer Narrative
The field report for a bent lead was confirmed.Analysis of the lead found a bent portion of the inner and outer coils at the distal region due to the implant procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TENDRIL STS
Type of Device
PERMANENT PACEMAKER ELECTRODE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
denise johnson
5050 nathan lane north
plymouth, MN 55442
6517564470
MDR Report Key6228855
MDR Text Key64380589
Report Number2017865-2017-00137
Device Sequence Number1
Product Code NVN
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P960013
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/21/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/06/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/31/2019
Device Model Number2088TC/52
Device Lot NumberA000027044
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/23/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/18/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-