ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE) ASSURITY RF DR; IMPLANTABLE PACEMAKER PULSE GENERATOR Back to Search Results
Model Number PM2240
Device Problems Loose or Intermittent Connection (1371); Pacing Problem (1439); Connection Problem (2900); Programming Issue (3014)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/30/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that an asymptomatic patient presented for a lead revision after a car accident.The device was restricted to wand only telemetry and programming was very slow.Set screw anomaly and high pacing thresholds were also noted.The physician elected to replace the device.The patient was stable before, during and after the procedure.
 
Manufacturer Narrative
The reported field event of no rf telemetry was not confirmed in the laboratory.The device was tested on the bench and no anomalies were found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ASSURITY RF DR
Type of Device
IMPLANTABLE PACEMAKER PULSE GENERATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
645 almanor avenue
sunnyvale CA 94085
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
645 almanor avenue
sunnyvale CA 94085
Manufacturer Contact
denise johnson
5050 nathan lane north
plymouth, MN 55442
6517564470
MDR Report Key6229357
MDR Text Key64427222
Report Number2938836-2017-01288
Device Sequence Number1
Product Code LWP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970013
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/30/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/06/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/31/2017
Device Model NumberPM2240
Device Lot NumberA000013984
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/22/2016
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/31/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/02/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age81 YR
-
-