ST. JUDE MEDICAL, INC.(CRM-SYLMAR) MERLIN PROGRAMMER; PACEMAKER PROGRAMMER Back to Search Results
Model Number 3650
Device Problem Improper Device Output (2953)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/21/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported through remote transmission that the device had reached eri even though a device check 7 months prior revealed the longevity to be 1.4 years.It was noted that the previous longevity of 1.4 years was an overestimation by the programmer due to an outdated algorithm at the time and that the algorithm was fixed with software updates.The patient presented in-clinic for a routine check-up and no symptoms were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MERLIN PROGRAMMER
Type of Device
PACEMAKER PROGRAMMER
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
denise johnson
5050 nathan lane north
plymouth, MN 55442
6517564470
MDR Report Key6236630
MDR Text Key64290111
Report Number2017865-2017-00290
Device Sequence Number1
Product Code KRG
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Nurse
Type of Report Initial
Report Date 11/21/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/10/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number3650
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/21/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age83 YR
-
-