Model Number 2088TC/58
Device Problems Break (1069); Kinked (1339); Over-Sensing (1438)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/10/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that episodes of lead noise were observed.Lead damage was suspected.No further information is available at this time.
 
Event Description
New information received states that the lead was explanted and replaced.After the lead was explanted an acute kink and clavicular crush were observed.The patient tolerated the procedure well.
 
Manufacturer Narrative
Correction needed for return date from (b)(6) 2017.Mp (b)(6) 2017.
 
Manufacturer Narrative
A partial lead was returned for analysis in 2 segments.The damage found was sustained during the surgical procedure.The portion of the lead that was returned was otherwise normal.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TENDRIL STS
Type of Device
PERMANENT PACEMAKER ELECTRODE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
645 almanor avenue
sunnyvale CA 94085
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
645 almanor avenue
sunnyvale CA 94085
Manufacturer Contact
denise johnson
5050 nathan lane north
plymouth, MN 55442
6517564470
MDR Report Key6239201
MDR Text Key64619986
Report Number2938836-2017-02310
Device Sequence Number1
Product Code NVN
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P960013
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date 08/20/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/10/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/30/2016
Device Model Number2088TC/58
Device Lot Number4249778
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/16/2017
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/26/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/15/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age56 YR
-
-