Operative notes were received indicating the device was interrogated and functioning properly at the time of surgery (b)(6) 2016 and the impedance was okay and the device was programmed to previous settings.It also stated that the tachycardia monitoring was interrogated.This was functioning on a setting of #1.The autostimulator was set at 2.5 level of stimulation".Later it was stated the device interrogated appropriately.Lead impedance is okay.However, on notes from a post-operative follow-up (b)(6) 2016, tachycardia monitoring did not pick up his heart rate at a setting of 1, 2, 3, 4, or 5.At the time of the surgical procedure cardiac monitoring/tachycardia monitoring was functioning on a setting of #1.Today tachycardia monitoring was not picking up his heart rate even at a setting of sensitivity of #5.The device was left set at #5." lead impedance was ok.Cardiac monitoring functioning was not detecting his heart rate at a setting of 1, 2, 3, 4, 5.It was left set at #5.
|