The patient was undergoing a thrombectomy procedure using a penumbra system aspiration pump max 110 v (pump max).Prior to the procedure, the patient had carotid stenosis on both sides of the internal carotid artery (ica), which was treated by placing a stent in the left ica.After the stenting procedure, thrombus was confirmed in the distal m1 segment of the ica.Thereafter, the physician attempted to aspirate the thrombus using the pump max and a penumbra system 5max ace reperfusion catheter (5max ace).The pump max was powered on with the green light illuminated and the cooling fan operating; however, no vacuum was produced.Therefore, the procedure was successfully completed using a syringe.There was no report of an adverse effect to the patient.
|
Results: the penumbra system aspiration pump max 110v (pump max) was opened by a penumbra investigator and corrosion was observed on the outlet cylinder piston crown.Conclusions: evaluation of the returned pump max confirmed that even though power was supplied to the pump max, the vacuum pump would not turn on.The pump housing was removed, and the vacuum pump was opened by penumbra engineers.It was observed that the piston crown in the outlet cylinder was corroded.The observed corrosion likely caused the piston to seize inside the cylinder.The root cause of this corrosion could not be determined.The 5max ace mentioned in the complaint was not returned for evaluation.Penumbra pumps are visually and functionally inspected during incoming quality inspection.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|