Follow up submission: no sample was received for evaluation.The device history record for the lot number reported was evaluated and no issues were observed.Two years of complaints were reviewed related to this issue, and no trend was identified.The needle is supplied by an external vendor.The supplier was issued a scar.The supplier stated that the detachment may be caused by an insufficient epoxy at the site of the needle and the hub.The supplier will further determine the root cause and applicable action plan related to this issue.
|