BIOMET ORTHOPEDICS INDIANA TOME SINGLE; GUIDE, SURGICAL Back to Search Results
Model Number N/A
Device Problem Device Packaging Compromised (2916)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Event Description
It was reported that a hair was found inside the packaging.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Manufacture date - november 2015.
 
Manufacturer Narrative
Complaint sample was evaluated and the reported event was confirmed.Visual inspection of the returned product identified a hair like fiber in the sterile packaging.Further analysis determined that the ftir spectrum is consistent with skin and therefore the fiber is most likely hair.Dhr was reviewed and no discrepancies were found.The root cause of the reported issue is attributed to a manufacturing issue, and the need for further corrective actions to prevent future recurrence is being further reviewed.A summary of the investigation has been sent to the complainant.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INDIANA TOME SINGLE
Type of Device
GUIDE, SURGICAL
Manufacturer (Section D)
BIOMET ORTHOPEDICS
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
BIOMET ORTHOPEDICS
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key6251653
MDR Text Key64898621
Report Number0001825034-2017-00170
Device Sequence Number1
Product Code FZX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PEXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,health profession
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 05/30/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/13/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number200060
Device Lot Number977750
Other Device ID NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/16/2017
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/25/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/17/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
-
-