The insulin pump was received with critical pump error and pump error 35 was present in long trace file due to corrosion on force sensor assembly.We were unable to perform displacement test, rewind test, prime/seating test, basic occlusion test, force sensor test, occlusion test, sleep current measurement, active current measurement and self-test due to critical pump errors.The insulin pump was received with scratched casing and minor scratches on lcd window.The insulin pump was received with pillowing keypad overlay, cracked case on battery tube area, peeling end cap address label, moisture damage on motor assembly and severe corrosion on all electronic assemblies.The insulin pump involved in this event is the 640g insulin infusion pump, which is not marketed in the united states.However, the device is similar to the paradigm real-time insulin infusion pump, which is marketed in the united states.
|