• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ANIMAS CORPORATION ANIMAS VIBE; INSULIN INFUSION PUMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANIMAS CORPORATION ANIMAS VIBE; INSULIN INFUSION PUMP Back to Search Results
Device Problem Temperature Problem (3022)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
The pump has not been returned to animas for evaluation.If the device is returned, an evaluation shall be completed and a supplemental report will be filed.No conclusions can be made at this time.
 
Event Description
On (b)(6) 2017, the reporter contacted animas, alleging a temperature (temp - no physical damage) issue.There was no indication that the product caused or contributed to an adverse event.This issue is being reported because there is the potential for the user to experience an injury to the skin due to increased pump temperature.
 
Manufacturer Narrative
The device has been returned and evaluated by product analysis on 03/07/2017 with the following findings: during investigation, there was no activity relating to the complaint found in the black box or download history.There was no damage found to the battery compartment or cap.There was no overheating during testing.The pump electrical current draws were found to be within specification.The pump was opened and there was no damage or evidence of internal moisture found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANIMAS VIBE
Type of Device
INSULIN INFUSION PUMP
Manufacturer (Section D)
ANIMAS CORPORATION
200 lawrence dr
west chester PA 19380 3428
Manufacturer (Section G)
ANIMAS CORPORATION
200 lawrence dr
west chester PA 19380 3428
Manufacturer Contact
karin sargrad
200 lawrence dr
west chester, PA 19380-3428
4843561808
MDR Report Key6255637
MDR Text Key65361580
Report Number2531779-2017-01320
Device Sequence Number1
Product Code MDS
UDI-Device Identifier10840406100051
UDI-Public0110840406100051
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P130007
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Patient
Type of Report Initial,Followup
Report Date 01/04/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/17/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Other Device ID Number1-MCBC-3762
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/18/2017
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? Yes
Device Age21 MO
Date Manufacturer Received01/04/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/27/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age68 YR
Patient Weight175
-
-