Model Number MMT-1712W
Device Problems Break (1069); Component Missing (2306); Naturally Worn (2988)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/25/2016
Event Type  malfunction  
Manufacturer Narrative
Analysis was unable to perform displacement test due to missing retainer.The insulin pump was received missing o-ring at the reservoir tube, broken reservoir tube lip, keypad overlay texture damage, cracked select button keypad overlay and minor scratches on lcd window.The insulin pump involved in this event is the 640g insulin infusion pump, which is not marketed in the united states.However, the device is similar to the paradigm real-time insulin infusion pump, which is marketed in the united states.Osed to the public under the freedom of information act.
 
Event Description
The customer reported via phone call that the retainer ring and o-rings were missing.The customer also reported that the display window was worn and damaged.The customer's blood glucose was unknown at the time of incident.The insulin pump was returned for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
640G INSULIN PUMP
Type of Device
INSULIN INFUSION PUMP / SENSOR AUGMENTED
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50
road 31 km 24.4
juncos PR 00777 3869
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50
road 31 km 24.4
juncos PR 00777 3869
Manufacturer Contact
gerwin de graaff
18000 devonshire street
northridge, CA 91325-1219
8185764805
MDR Report Key6257008
MDR Text Key65357041
Report Number3004209178-2017-84162
Device Sequence Number1
Product Code OYC
Combination Product (y/n)N
Reporter Country CodeFR
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial
Report Date 10/25/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/17/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMMT-1712W
Device Catalogue NumberMMT-1712W
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/09/2016
Is the Reporter a Health Professional? No
Date Manufacturer Received10/25/2016
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-