MAUDE Adverse Event Report: BAXTER HEALTHCARE - LARGO HOMECHOICE PRO; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE - LARGO HOMECHOICE PRO; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY Back to Search Results
Catalog Number 5C8310
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Complaint, Ill-Defined (2331)
Event Type  Injury  
Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
A peritoneal dialysis (pd) patient experienced an unspecified sentinel event while connected to the homechoice device.The event occurred while the patient was hospitalized for an unspecified condition.The sentinel event was not further described.The cause of the event, treatment details, action taken with pd therapy, and the patient¿s recovery status were not reported.No additional information is available.
 
Manufacturer Narrative
The device was received for evaluation.The event history log review showed no keystrokes, programming, or use related events that indicated and/or contributed to the reported issue.A service history review revealed no indication that the parts replaced during servicing caused or contributed to the reported event.An internal and external inspection was performed and no issues were noted.The homechoice device received a returned instrument testing evaluation (rite).This evaluation included functional and electrical testing of the device.The device was determined to meet functional and electrical performance specification requirements per rite testing.The reported issue could not be verified through the device evaluation.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HOMECHOICE PRO
Type of Device
SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
Manufacturer (Section D)
BAXTER HEALTHCARE - LARGO
largo FL
Manufacturer (Section G)
BAXTER HEALTHCARE - LARGO
7511 114th ave. north
ni
largo FL 33773
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key6265458
MDR Text Key65344015
Report Number1416980-2017-00298
Device Sequence Number1
Product Code FKX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102936
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 03/06/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/19/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Catalogue Number5C8310
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/22/2016
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/03/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Treatment
UNSPECIFIED DIALYSIS SOLUTIONS
Patient Outcome(s) Hospitalization;
-
-