• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: GIVEN LTD (ISRAEL) BRAVO PH RECORDER REFURBISHED; ELECTRODE, PH, STOMACH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GIVEN LTD (ISRAEL) BRAVO PH RECORDER REFURBISHED; ELECTRODE, PH, STOMACH Back to Search Results
Model Number RFB-0450
Device Problem Calibration Problem (2890)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/19/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
According to the reporter they have a bravo recorder which has a mismatch error.The capsule was in a patient when this issue was noticed.There was no injury to the patient or user, and a repeat procedure will be required.
 
Manufacturer Narrative
Evaluation summary: one recorder was received for evaluation and was inspected for external damage.According to the reported information, the customer successfully calibrated the recorder with the capsule, but then mistakenly entered the calibration menu instead of "start study", then selected "calibrate capsule" a second time, followed by "record study" which led to the error of "id mismatch".The product performed to specification and the issue is attributed to user mishandling.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BRAVO PH RECORDER REFURBISHED
Type of Device
ELECTRODE, PH, STOMACH
Manufacturer (Section D)
GIVEN LTD (ISRAEL)
2 hacarmel st.
new industrial pob 258,
yoqneam
IS 
Manufacturer (Section G)
GIVEN LTD (ISRAEL)
2 hacarmel st.
new industrial pob 258,
yoqneam
IS  
Manufacturer Contact
amy beeman
5920 longbow drive
boulder, CO 80301
7632104064
MDR Report Key6269648
MDR Text Key65508148
Report Number9710107-2017-00040
Device Sequence Number1
Product Code FFT
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102543
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 04/11/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/23/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberRFB-0450
Device Catalogue NumberRFB-0450
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/18/2017
Is the Reporter a Health Professional? No
Date Manufacturer Received01/04/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Weight59
-
-