MAUDE Adverse Event Report: MAKO SURGICAL CORP. TIBIAL INSERT ONLAY TRIAL-SIZE 4-8MM; PROSTHESIS, KNEE PATELLOFEMOROTIBIAL, PARTIAL, SEMI-CONSTRAINED, CEMENTED, POLYM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAKO SURGICAL CORP. TIBIAL INSERT ONLAY TRIAL-SIZE 4-8MM; PROSTHESIS, KNEE PATELLOFEMOROTIBIAL, PARTIAL, SEMI-CONSTRAINED, CEMENTED, POLYM Back to Search Results
Catalog Number 170704-1
Device Problems Material Deformation (2976); Material Integrity Problem (2978); Insufficient Information (3190)
Patient Problems No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Event Date 01/03/2017
Event Type  malfunction  
Manufacturer Narrative
A supplemental report will be submitted if additional information becomes available.
 
Event Description
It was reported that parts of poly insert were chipping off.
 
Manufacturer Narrative
An event regarding damage involving an mako tibial insert was reported.The event was confirmed.Method and results: device evaluation and results: visual inspection of returned device has been done and device found damaged.Examination of the returned device with material analysis engineer indicated damage observed are explantation damage.Medical records received and evaluation: no medical records or x-rays were made available for evaluation.Device history review: device history review indicated the devices accepted into final stock from the reported lot were free from discrepancies.Complaint history review: there has been no other event for the lot referenced.Conclusions: the visual inspection confirms the event as the returned product was found damaged.Moreover the examination of the returned device with material analysis engineer indicated damage observed are explantation damage.No further investigation is possible at this time.If additional information is received, this investigation will be reopened and re-evaluated.
 
Event Description
It was reported that parts of poly insert were chipping off.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TIBIAL INSERT ONLAY TRIAL-SIZE 4-8MM
Type of Device
PROSTHESIS, KNEE PATELLOFEMOROTIBIAL, PARTIAL, SEMI-CONSTRAINED, CEMENTED, POLYM
Manufacturer (Section D)
MAKO SURGICAL CORP.
2555 davie road
fort lauderdale FL 33317
Manufacturer (Section G)
MAKO SURGICAL CORP.
2555 davie road
fort lauderdale FL 33317
Manufacturer Contact
keyla colon
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key6269783
MDR Text Key65527784
Report Number3005985723-2017-00033
Device Sequence Number1
Product Code NPJ
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 04/25/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/23/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number170704-1
Device Lot Number12021115
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/10/2017
Is the Reporter a Health Professional? No
Date Manufacturer Received03/28/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/17/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-