During preparation for a thrombectomy procedure, the penumbra system ace 68 reperfusion catheter broke into two pieces in the middle of the catheter upon removal from the packaging.The damage to the ace 68 occurred prior to use and therefore, it was not used in the procedure.The procedure was continued using a penumbra system ace 60 reperfusion catheter (ace 60).
|
Please note that the device is no longer available for return as mentioned in the initial mdr; therefore, the mdr was updated accordingly.The product was not returned for evaluation.Without the return of the device, the root cause of the problem cannot be determined.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|