Boston scientific received information that the patient with this right ventricular (rv) defibrillation lead presented to the emergency room (er) due to heating/warmth sensations in the throat and pocket during pacing.The shock impedance measurements had been trending up, and were subsequently greater than 200 ohms.The patient also had flushing on the left-sided neck area.Their neck was distended, and the patient had pulsations and tremors on the left pectoral, which could not be recreated upon troubleshooting.The patient also reported feeling sluggish.Lead diagnostics were performed and the patient felt every ventricular pace.The rv pacing impedance measurement was 682 ohms and consistent.A surgical revision was performed; the lead was partially abandoned, and a portion of the lead was returned for analysis.A new system was implanted on the patient¿s right side.It was suspected that the patient had subclavian syndrome.Additional information was received which indicated that during the revision, this rv lead was tested via the pacing system analyzer (psa) and still showed high shock impedance measurements.There was no visible injury to the rv lead.The physician chose to replace the lead and utilize the same device.The system was tested during the pre-discharge check and was noted to be fine.No additional adverse patient effects were reported.
|